●In line with the requirements of the EU EN12469 standard for Class II biological safety cabinets to be classified as Type A2, with corresponding TUV/CE certificates.
●Vertical laminar negative pressure model. 70% of the air flow is filtered and recycled, and 30% of the air flow can be discharged into the room or connected to the exhaust system after being filtered.
●High-definition LCD human-machine dialogue interface, touch key operation, real-time display of wind speed, filter operation status, and provide fault alarm.
●The imported intelligent air volume automatic compensation system ensures that the fan air volume changes less than 10% when the filter resistance increases by 50%, improving safety.
●Each biological safety cabinet has passed the factory test of safety and product performance in accordance with international standards.
●In line with the requirements of the EU EN12469 standard for Class II biological safety cabinets to be classified as Type A2, with corresponding TUV/CE certificates.
●Vertical laminar negative pressure model. 70% of the air flow is filtered and recycled, and 30% of the air flow can be discharged into the room or connected to the exhaust system after being filtered.
●High-definition LCD human-machine dialogue interface, touch key operation, real-time display of wind speed, filter operation status, and provide fault alarm.
●The imported intelligent air volume automatic compensation system ensures that the fan air volume changes less than 10% when the filter resistance increases by 50%, improving safety.
●Each biological safety cabinet has passed the factory test of safety and product performance in accordance with international standards.
Friendly Control Panel ◆ LED screen, show the air velocity, time, fan station, alarm in real time.
|
|
Safe and Reliable, Energy-Efficient ◆ DC fan and LED light make the operation power consumption save 30%. ◆ Intelligent auto-supply system of imported blower ensures the change of air volume less than 10% when the resistance of filter increases 50% and enhance safety. ◆ The electrical parts are certified by VDE and spare sockets are accord with European standard. ◆ The alarm system for safe operation height is interlocked with light. ◆ Leak detection in cabinet, no leakage at 500Pa.
Humanized design ◆ There is key for power switch, so you can make management for operators. ◆ 10°sloped front make it comfortable to operator. ◆ Negative pressure and double layers chamber around ensure non-leakage, non-pollution. Demountable circular-arc armrest can alleviate working tiredness and make it easy to transport. ◆ Front window adopts hanging lifting system (Imported from Japan) and more than 5mm thickness security glass with arbitrary positioning, high stability and convenient to sterilization when it closed. ◆ There are two anti-splash spare sockets in the cabinet. ◆ Movable stainless steel worktable 4 liter capacity stainless steel collecting tank with below down valve, easy to collect water, cleaning and disinfection. ◆ Non-obstructive air return mode in the front suction channel. ◆ The adjustable leveling foot is without exposure thread.
|
|
HEPA filters ◆ USA HV HEPA Filter Medium, provide 99.995% typical efficiency for particle size over 0.3 microns, and the materials is glass fiber. ◆ A minimum ISO 5 air quality in the work area ◆ Unique Technology of non leakage. ◆ H14 HEPA filters are accord with EN13091:1999 and EN 1822-1 standard. Every HEPA filters will be tested to ensure non leakage. |
|
Model/Item | BSC-1003ⅡA2 | BSC-1303ⅡA2 | BSC-1603ⅡA2 | |
Nominal Size | 1meter(3') | 1.3meter(4') | 1.6meter(5') | |
Filtration Efficiency | HEPA: ≥99.995%, @0.3μm / ULPA: ≥99.9995%, @0.12μm | |||
Downflow Velocity | 0.35 m/s(68.88fpm) | |||
Inflow Velocity | 0.55 m/s(108.24fpm) | |||
Noise | ≤65 dB(A) | |||
| Vibration | ≤5 μm(center of tabletop) | |||
Power Supply | AC220~240V,1φ,50/60Hz | |||
| Opening Height | 200 mm | |||
Operator Protection | A. Total colony in impaction sampler ≤10CFU./time B. Total colony in slot sampler ≤5CFU./time | |||
Product Protection | Total colony in culture dish ≤5CFU./time | |||
Cross Contamination | Total colony in culture dish ≤2CFU./time | |||
Max Consumption (with spare socket) | 1.5 kVA | 1.5 kVA | 1.8 kVA | |
Rated Power (without spare socket) | 0.5 kvA | 0.54 kVA | 0.75 kVA | |
Weight | 280 kg | 300 kg | 320 kg | |
Work Dimension (W1XD1XH1) |
970*600*620 mm (38.2''x23.6''x24.4'')
| 1270*600*620 mm (50''x23.6''x24.4'') | 1570*600*620 mm (61.8''x23.6''x24.4'') | |
Overall Dimension (WXDXH)
| 1200*790*2050 mm (47.2''x31.1''x80.7'') | 1500*790*2050 mm (59.1''x31.1''x80.7'') | 1800*790*2050 mm (70.9''x31.1''x80.7'') | |
Size (W) and Qty. of Light | LED 24.5 w x ① | LED 31 w x ① | LED 36 w x ① | |
Size (W) and Qty. of Light on-duty | 18 w x ① | 30 w x ① | 36 w x ① | |
Lumin. | ≥750 Lx | |||
| Cabinet Materials | High-grade steel and lacquered in ivory | |||
| Working Area Materials | SS304 fully finished | |||
Air Direction | Top out | |||
| Plug Style | Europe Type(AC 250V/10A) | |||
| Qty. of Spare Socket | Europe Type(AC 250V/2A) 2 units | |||
| No. | Name | Note |
| 1 | Electric Socket | Standard, 2 pcs |
| 2 | U.V.Light | Optional |
| 3 | Canopy Connecter | Optional |
| 4 | Gas Valve | Optional |
| 5 | Vacuum Valve | Optional |
| 6 | Water Valve | Optional |
| 7 | I.V.Pole | Optional |